Explore TPC-Port Neches' site monitoring solutions designed to ensure cleaner air and a healthier environment.
TPC conducts continuous ambient air monitoring using Gas Chromatographs (GCs) with data available to the facility and the public. The GCs continuously acquire an ambient air sample and through an analysis, use chromatography to separate and selectively analyze for the target analytes isobutylene and 1,3-butadiene. The GCs provide one data point for 1,3-butadiene and isobutylene in 15-minute increments, these data are reported in units of parts per billion (ppb) and represent the 15-minute sample average. The reported ppb concentration units are representative of the ambient air concentrations for each pollutant.
Continuous ambient air monitoring is conducted at the facility, including monitoring for analyte concentrations of 1,3-butadiene and isobutylene and meteorological parameters wind direction, wind speed, temperature, relative humidity and barometric pressure.
GC data are reported in 15-minute intervals and paired with meteorological data, these data are transferred automatically using remote transmission. Each data point represents a 15-minute average concentration where the time stamp is the end of the averaging period. Daily one-point QC checks are conducted to verify instrument performance.
Quality assurance is a fundamental part of the air monitoring program, following established procedures and approved methods to ensure data is accurate and reliable. Part of quality assurance includes routine quality checks such as daily QC checks, quarterly multipoint verifications and audits outlined below.
Daily one-point QC checks are conducted on each gas chromatograph. Each check includes two QC checks and two zero checks.
QC checks should be within 30% of target values. Zero checks should be below 5 ppb for isobutylene and 0.5 ppb for 1,3-butadiene. If the checks fall outside these limits, operators will be prompted to investigate and, based on findings, data may be invalidated as needed.
Invalid flags or dashes "--" are reported for the data if it is not valid.
The leading edge of GC data (generally up to 24 hours) is considered preliminary as it is not bracketed by a second QC check. The data is auto validated when the next daily QC check is performed, but is subject to change pending operator investigation and review. The data will undergo a rigorous review and reporting process quarterly, after which the data can be considered certified.
The method detection limit or MDL is the lower limit of instrument response where the target analytes are distinguishable from blanks. These values are determined through experience and experimentation. GC values below the MDL are displayed as "<MDL".
Multipoint verifications (MVER) are completed on GCs quarterly or as needed. These checks verify sensor response across the operating range and give operators the opportunity to adjust sensor response if needed.
Performance audits are conducted on the GCs quarterly or as needed. These checks verify sensor response across the sensor operating range using third-party flow meters, gas standards and an operator not typically involved in day-to-day operation. Audit findings are communicated to the operators and corrective action is taken if needed.
Petroleum refineries and petrochemical plants
Cigarette smoke
Natural sources (forest fires)
Vehicle emissions (especially in areas with heavy traffic)
Chemical production facilities and synthetic rubber manufacturing plants
Petroleum refineries
Vehicle emissions (especially in areas with heavy traffic)
Fuel combustion processes
Manufacturing plants (chemical, rubber, polymer production)
Natural gas processing
Industrial solvents